GW Instek GPM-8340 - Introduction of GPM-8340 4-Element AC Power Meter
▪ Introduction of GPM-8340 4-Element AC Power Meter
The GW Instek GPM-8340 is a digital power meter designed specifically for multiphase AC power measurements, making it suitable for a wide range of electrical and electronic product testing applications. Featuring a test bandwidth of DC, 0.1 Hz to 10 kHz, a 16-bit A/D converter, and a 300 kHz sampling rate, the GPM-8340 utilizes a 5” TFT LCD display, 5-digit measurement display, provides 28 relevant power measurement parameters, and offers high-precision measurement capabilities. It also features waveform display (voltage/current/power), integration measurement function, harmonic measurement and analysis (50 Hz / 60Hz measurement compliant with IEC61000-4-7 requirements), external sensor input terminals, and various communication interfaces, helping users achieve clear, convenient, and accurate power measurements. It is considered one of the most feature-rich and cost-effective power meters in its class.
Product Features
- ※ 4 Input Elements
- ※ 5” TFT LCD
- ※ Voltage/Current Test Freq. Bandwidth: DC, 0.1 Hz to 10 kHz
- ※ Lowest current range 5 mA & 0.1 µW watt resolution
- ※ Waveform Display: V (Voltage), I (Current), P (Wattage)
- ※ Meeting IEC 61000-4-7 harmonic measurement (50 Hz/ 60 Hz)
- ※ Wiring selecting button (1P2W, 1P3W, 3P3W, 3P4W, 3V3A)
- ※ Harmonic measurement & analysis up to 50 orders
- ※ Supports inrush current and energy measurement
- ※ Provide external current sensor input (EXT)
- ※ Standard Interface: RS-232C, USB device/ host, LAN, Digital I/O and mini GPIB or CAN (select one)
| GENERAL | |||
|---|---|---|---|
| ⚠ Note | The below are the basic conditions required to operate the GPM-8340 within specifications:
Calibration: Yearly
Operating environment: (18 to 28) °C (64.4 to 82.4) °F Humidity: < 80 % RH, Accuracy: ± (% of reading + % of range) The specifications apply when it warmed up for at least 30 min and operates in the slow rate. The power supply cable must be grounded to ensure accuracy. Input voltage and current must be standard sine wave. The power factor must be 1. The common-mode voltage must be zero. |
||
| Specification Condition | Temperature: 23 °C ± 5 °C; Humidity: < 80 %RH (non-condensing) | ||
| Operation Condition | Temperature 0 °C to 40 °C 30 °C to 40 °C, Relative Humidity < 70 %RH (non-condensing) > 40 °C, Relative Humidity < 50 %RH (non-condensing) Indoor use only Altitude: < 2000 m Pollution degree 2 |
||
| Storage Condition | Temperature −10 °C to 70 °C; Humidity: < 90 %RH (non-condensing) | ||
| Power Source | AC 100 V to 240 V, 50 Hz to 60 Hz ; Consumption Max. 140 VA | ||
| Dimensions | 220 mm(W) x 132 mm(H) x 444 mm(D) | ||
| Weight | Approx. 5.24 kg | ||
| INPUT | |||
| Item | Specifications | ||
| Input Type | Voltage | Floating input through resistive voltage divider | |
| Current | Floating input through shunt | ||
| Measure Range | Voltage (CF=3) | 15 V, 30 V, 60 V, 150 V, 300 V, 600 V and 1000 V | |
| Current (CF=4) | |||
| Direct input | 5 mA, 20 mA, 50 mA, 200 mA, 500 mA, 2 A, 5 A, 20 A | ||
| Sensor input | EXT: 10 mV, 25 mV, 50 mV, 100 mV, 2.5 V, 5 V, 10 V | ||
| Input Impedance | Voltage | Input resistance: approach 4 MΩ | |
| Current | |||
| Direct input range (5 to 200) mA | Input resistance: approach 505 mΩ | ||
| Direct input range (0.5 to 20) A | Input resistance: approach 5 mΩ | ||
| Sensor input | Input resistance: approach 100 kΩ | ||
| Continuous Maximum Allowable Input |
Voltage | Peak value of 1.5 kV or RMS value of 1 kV, whichever is less. | |
| Current | |||
| Direct input range (5 to 200) mA | Peak value of 30 A or RMS value of 20 A, whichever is less. | ||
| Direct input range (0.5 to 20) A | Peak value of 100 A or RMS value of 30 A, whichever is | ||
| Sensor input | Peak value less than or equal to 5 times of the rated range | ||
| Input Bandwidth | DC, 0.1 Hz to 10 kHz | ||
| Continuous Maximum Common-mode Voltage |
600 Vrms, CAT Ⅱ | ||
| Line Filter | select OFF or ON (cut off frequency of 5.5 kHz) | ||
| Frequency Filter | select OFF or ON (cut off frequency of 5.5 kHz) | ||
| A/D Converter | Simultaneous conversion voltage and current inputs | ||
| Resolution 16 bits | |||
| Maximum conversion rate Approx. 300 kHz | |||
| VOLTAGE AND CURRENT ACCURACY | |||
| Item | Specifications | ||
| Requirements | Temperature | 23 °C ± 5 °C | |
| Humidity | 30 % to 75 %RH | ||
| Input waveform | Sine wave | ||
| common-mode voltage | 0 V | ||
| Number of displayed digits | 5 digits | ||
| Frequency filter | Turn on to measure voltage or current of 200 Hz or less | ||
| After 30 minutes after warm-up time has passed | |||
| After measurement range is changed (zero-level compensation) | |||
| Update interval is 250 ms | |||
| Accuracy | Effective range | 1 % to 105 % of range | |
| DC | ± (0.1 % of reading + 0.2 % of range) | ||
| 0.1 Hz ≤ f < 45 Hz | ± (0.1 % of reading + 0.2 % of range) | ||
| 45 Hz ≤ f ≤ 66 Hz | ± (0.1 % of reading + 0.05 % of range) | ||
| 66 Hz < f ≤ 1 kHz | ± (0.1 % of reading + 0.2 % of range) | ||
| 1 kHz < f ≤ 10 kHz | ± ((0.07 x f) % of reading + 0.3% of range) | ||
| Temperature Coefficient | Add | ± 0.03 % of reading/°C within the range 5 °C to 18 °C or 28 °C to 40 °C. | |
| When the Line Filter is Turned ON |
45 Hz to 66 Hz | Add 0.2 % of reading | |
| < 45 Hz | Add 0.5 % of reading | ||
| Accuracy Changes Caused by Data Update Interval | When the data update interval is 100 ms, and Auto, add 0.05 % of reading to the 0.1 Hz to 1 kHz accuracy. | ||
| Influence of Temperature Changes After Zero-level Compensation or Range Change | Add 0.02 % of range/°C to the DC voltage accuracy. | |
|---|---|---|
| Add the following value to the DC current accuracies. | ||
| 5 mA / 20 mA / 50 mA / 200 mA ranges | 5 μA/°C | |
| 0.5 A / 2 A / 5 A / 20 A ranges | 500 μA/°C | |
| EXT: 2.5 V/ 5 V/ 10 V | 1 mV/°C | |
| EXT: 10 mV/ 25 mV/ 50 mV/ 100 mV | 50 μV/°C | |
| ACTIVE POWER ACCURACY | ||
| Item | Specifications | |
| Requirements | Same as the conditions for voltage and current. | |
| Power factor | 1 | |
| Accuracy | Effective range | 1 % to 110 % of range |
| DC | ± (0.1 % of reading + 0.2 % of range) | |
| 0.1 Hz ≤ f < 45 Hz | ± (0.3 % of reading + 0.2 % of range) | |
| 45 Hz ≤ f ≤ 66 Hz | ± (0.1 % of reading + 0.05 % of range) | |
| 66 Hz < f ≤ 1 kHz | ± (0.2 % of reading + 0.2 % of range) | |
| 1 kHz < f ≤ 10 kHz | ± (0.1 % of reading + 0.3 % of range) ± [{0.067 x (f-1)} % of reading] | |
| Influence of Power Factor | when power factor (λ) = 0 (S: apparent power) | |
| ± 0.1 % of S for 45 Hz ≤ f ≤ 66 Hz | ||
| ± {(0.1 + 0.15 × f) % of S } | ||
| •f is frequency of input signal in kHz | ||
| when 0 < λ < 1 (Φ: phase angle of the Voltage and current) | ||
| (power reading ) × [(power reading error%) + (power range %) × (power range / indicated | ||
| When the Line Filter is Turned ON |
45 Hz to 66 Hz | Add 0.3 % of reading |
| < 45 Hz | Add 1 % of reading | |
| Temperature Coefficient | Same as the temperature coefficient for voltage and current | |
| Accuracy of Apparent Power S | Voltage accuracy + current accuracy | |
| Accuracy of Reactive Power Q | Accuracy of apparent power + (√1.0004 - λ2) - (√1 - λ2) × 100 % | |
| Accuracy of Power Factor λ | ± [(λ-λ/1.0002)+ | cosø-cos{ø+sin-1 (influence from the power factor when λ = 0 %/100)} | ] ± 1 digit when voltage and current are at the measurement range rated input |
|
| Accuracy of Phase Difference Φ | ± [ | ø-cos-1(λ/1.0002) | + sin-1 (influence from the power factor when λ = 0 % / 100)] ± 1 digit when voltage and current are at the measurement range rated input | |
| VOLTAGE, CURRENT AND ACTIVE POWER MEASUREMENTS | ||
| Item | Specifications | |
| Measurement Method | Digital sampling method | |
| Crest Factor | 3 (voltage) or 4 (current) | |
| Wiring System | 1P2W, 1P3W, 3P3W, 3P4W, 3V3A | |
| Peak Measurement | Select manual or auto ranging | |
| Auto Range | Auto-range increase | |
| The range is upped when any of the following conditions is met. | ||
| Vrms or Irms exceeds 130 % of the currently set measurement range. Vpk, Ipk value of the input signal exceeds 300 % of the currently set measurement range. |
||
| Auto-range decline | ||
| The range is downed when all of the following conditions are met. | ||
| Vrms or Irms is less than or equal to 30 % of the measurement range. Vrms or Irms is less than or equal to 125 % of the next lower measurement range. Vpk, Ipk value of the input signal exceeds 300 % of the currently set measurement range. |
||
| Display Mode Switching | AC+DC (the true RMS value of voltage and current) | |
| V-MEAN (the rectified mean value calibrated to the RMS value of the voltage and the true RMS value of the | ||
| AC | ||
| DC | ||
| Measurement Synchronization Source | Select voltage, current, or off | |
| In the case of Auto Update Rate, select the voltage or current from the equipped element. | ||
| Line Filter | Select OFF or ON (cutoff frequency at 5.5 kHz). | |
| Peak Measurement | Measures the peak (max, min) value of voltage, current or power from the instantaneous voltage, instantaneous current or instantaneous power that is sampled. | |
| Zero-level Compensation | Removes the internal offset of the measure unit (After measurement range is changed) | |
| Measurement Parameters | Voltage | Vrms , Vmn, Vdc , Vac |
| Current | Irms , Idc , Iac | |
| Active Power | P | |
| Apparent Power | VA | |
| Reactive power | VAR | |
| Power Factor | PF | |
| Crest Factor | CFI, CFV | |
| Phase Angle | DEG | |
| Frequency | IHz and VHz | |
| Voltage Peak | V+pk and V-pk | |
| Current Peak | I+pk and I-pk | |
| Active Power Peak | P+pk and P-pk | |
| Total Harmonic Distortion | THDI and THDV | |
| Mathematical Computation | MATH, EFFi | |
| Maximum Current Ratio | MCR | |
| FREQUENCY MEASUREMENT | ||||
|---|---|---|---|---|
| Item | Specifications | |||
| Measurement Item | Voltage and current | |||
| Measurement Frequency Range | Data update interval | Measurement Frequency Range | ||
| 0.25 s | 10 Hz ≤ f ≤ 10 kHz | |||
| 0.5 s | 5.0 Hz ≤ f ≤ 10 kHz | |||
| 1 s | 2.0 Hz ≤ f ≤ 10 kHz | |||
| 2 s | 1.0 Hz ≤ f ≤ 10 kHz | |||
| 5 s | 0.5 Hz ≤ f ≤ 10 kHz | |||
| 10 s | 0.2 Hz ≤ f ≤ 10 kHz | |||
| 20 s | 0.1 Hz ≤ f ≤ 10 kHz | |||
| Auto ( * ) | 0.1 Hz ≤ f ≤ 10 kHz | |||
| ( * ) Limit of the measurement lower limit frequency by the Timeout setting | ||||
| Timeout | lower limit frequency | |||
| 1 s | 2.0 Hz | |||
| 5 s | 0.5 Hz | |||
| 10 s | 0.2 Hz | |||
| 20 s | 0.1 Hz | |||
| Measurement Range | Auto switching among six types: 100 mHz, 1 Hz, 10 Hz, 100 Hz, 1 kHz, 10 kHz. | |||
| Frequency Filter | Select OFF or ON (cut off frequency of 5.5 kHz) | |||
| Accuracy | Requirements | When the input signal level is 30 % or more of the measurement range. • Frequency filter is ON when measuring voltage or current of 200 Hz or less. |
||
| ± (0.06 % of reading) | ||||
| INTEGRATION | ||||
| Item | Specifications | |||
| Mode | Select manual integration mode, standard integration mode, or repetitive integration mode. | |||
| Timer | Automatically stop integration by setting a timer. | |||
| Selectable range: 0 hours 00 minutes 00 seconds to 9999 hours 59 minutes 59 seconds | ||||
| Count Overflow | WP: 999999 MWh / -99999 MWh | |||
| Accuracy | ± (Power accuracy (or current accuracy) + 0.1 % of reading) (fixed range) | |||
| Range Setting | Auto range or fixed range is available for Integration | |||
| Timer Accuracy | ± 0.02 % | |||
| Remote Control | Start, stop and reset operations are available using an external remote signal. | |||
| HARMONIC MEASUREMENT | ||||
| Item | Specifications | |||
| Measured Item | Voltage, Current, Power | |||
| Measured Method | Zero-cross simultaneous calculation method | |||
| Frequency Range | 10 Hz to 1.2 kHz. | |||
| FFT Data Length | 4096 | |||
| (Auto switch when both (50 / 60) Hz and update rate must be greater than or equal to 0.25 s) | ||||
| Sample Rate, Window Width, and Upper Limit of Analysis Orders* |
Fundamental Frequency | Sample rate | Window Width | upper limit of Analysis orders |
| 45 Hz to 55 Hz | f × 512 | 10 | 50 | |
| 54 Hz to 66 Hz | f x 512 | 12 | 50 | |
| FFT data length | 1024 | |||
| Sample Rate, Window Width, and Upper limit of Analysis Orders* |
Fundamental Frequency | Sample rate | Window Width | upper limit of Analysis orders |
| 10 Hz to 67 Hz | f × 1024 | 1 | 50 | |
| 67 Hz to 150 Hz | f × 512 | 2 | 32 | |
| 150 Hz to 300 Hz | f × 256 | 4 | 16 | |
| 300 Hz to 600 Hz | f × 128 | 8 | 8 | |
| 600 Hz to 1200 Hz | f × 64 | 16 | 4 | |
| Accuracy | Frequency | Voltage | Current | Power |
| 10 Hz ≤ f < 45 Hz | 0.15 % of reading | 0.15 % of reading | 0.35 % of reading | |
| + 0.35 % of range | + 0.35 % of range | + 0.50 % of range | ||
| 45 Hz ≤ f < 440 Hz | 0.15 % of reading | 0.15 % of reading | 0.25 % of reading | |
| + 0.35 % of range | + 0.35 % of range | + 0.50 % of range | ||
| 440 Hz ≤ f < 1.2 kHz | 0.20 % of reading | 0.20 % of reading | 0.40 % of reading | |
| + 0.35 % of range | + 0.35 % of range | + 0.50 % of range | ||
| * 50 Hz / 60 Hz Compliant IEC61000-4-7 (update rate must be > 0.25 s). * Harmonic calculation: FFT method in which FFT data length is divided into 2 types: 1024 and 4096. * FFT data length automatically switches in accord with the Frequency and Update Rate of measured signal. |
||||
| REMOTE CONTROL INPUT/OUTPUT SIGNAL | |
|---|---|
| Item | Specifications |
| Remote Control Input Signal | Is_trig, Limit_trig |
| I/O Level | TTL |
| I/O Logic Format | Negative logic, Falling edge |
| Remote Control Out Ssignal | M1_Pass, M2_Pass, M3_Pass, M4_Pass |
| I/O Level | Relay form A |
| Output Rate Voltage | 30 Vdc |
| I/O Sink Current | Max. 0.5 A |
| * Q (VAR), S (VA), λ (PF) and Φ (DEG) are originated from the measured values including voltage, current and active power which go through computation process. In respect to distorted signal input, accordingly, the value acquired from other instruments, which employ different methods, may differ from that acquired from GPM-8340 unit. * “Zero” will be shown for S or Q and “--“ will be displayed for λ and Φ when either current or voltage is less than 0.5 % of the rated range. |
|
Specifications subject to change without notice. PM-8340CD1BH